Electromagne c Immunity (con nued):

Guidance and Manufacture's Declara on: Electromagne c Immunity
SAFIRA is intended for use in the electromagne c environment specified below.
The customer and/or user of SAFIRA should assure that it is used in such an environment.
Immunity
IEC 60601 Test
Test
Level
Conducted
3V rms (1kHz
RF IEC 61000
80%) 150kHz –
-4-6
80MHz
6V rms ISM and
amateur radio
bands
Radiated RF
3V/m (1kHz
IEC 61000-4-
80%) 80MHz –
3
2.7GHz
Note 1: The guidelines may not apply in all situa ons, Electromagne c propaga on is
affected by absorp on and reflec on from structures, objects and people.
MP-0007-3001 v4.0
Compliance
Electromagne c Environment: Guidance
Level
3Vms
Portable and mobile RF communica on
equipment should be no closer to any part of
SAFIRA, including cables, than the
recommended separa on distance of 300mm.
6V rms
Field strength from fixed RF transmi ers, as
determined by an electromagne c site
a
survey
, should be less than the compliance
level in each frequency range
3V/m
Interference may occur in the vicinity of
equipment marked with the following symbol:
b
.