- Automatic cleaning
Use an ISO 15883 compliant washer-disinfector that uses an alkaline
cleaning agent with a maximum pH of up to 10.
- Disinfection
External manual: disinfect using a clean cloth soaked in a suitable
substance (e.g. O-phenylphenol or ethyl alcohol). Products not
recommended: disinfectants based on benzalkonium chloride, acetone
or glutaraldehyde.
Internal manual: Using a suitable spray, follow the corresponding
instructions and spray the substance inside the head. Immediately after
internal disinfection, perform the oil lubrication process and / or the
sterilisation process.
Automated: Use an ISO 15883 compliant washer-disinfector that uses
an alkaline cleaning agent with a maximum pH of up to 10.
62
Eng