Immunity Test
Immunity test
Electrostatic
Contact ±8 kV
discharge (ESD)
IEC 61000-4-2
Air ±15 kV
Electrical fast
5/50 ns,
transient/burst
100 kHz, ±2 kV
IEC 61000-4-4
10. H O T A I R G E N E R AT O R
IEC 60601
test level
Contact ±8 kV
Air ±15 kV
5/50 ns,
100 kHz, ±2 kV
Compliance
level
environment – Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be similar to
that of a typical residen-
tial, commercial or
hospital environment. If
the user requires
continuous function, it is
recommended that the
VOLCANO MEDIC 2
Vaporization System is
operated from an
uninterruptible power
supply or battery.
125
Electromagnetic