Starting the 2100 Bioanalyzer Instrument

Turning on the 2100 Bioanalyzer Instrument

Turning on the 2100 Bioanalyzer Instrument
Condensation within the instrument
C AUTI ON
Temperature fluctuations could cause condensation within the instrument.
1 Turn on your Agilent 2100 Bioanalyzer instrument. The power switch is
2 The status indicator lamp will light green if power is present and all
Status Indicator
Green - Ready
Green, flashing - Busy
Orange - Self-test
Red - Error
126
Do not use the instrument under conditions with temperature fluctuations.
Let the instrument equilibrate to room temperature for one day to avoid
damage of the electronics.
located at the rear of the instrument, where the power cable plugs in (see
step 4 on page 123).
instrument self-tests have passed successfully.
If the status indicator lamp lights red, this indicates a hardware error. In this
case, please refer to the Troubleshooting section of the Agilent 2100
Bioanalyzer Maintenance and Troubleshooting Guide or the Online Help.