Descargar Imprimir esta página

Alarms And Troubleshooting - W&H Sterilizer Calibration Device Manual Del Usuario

Publicidad

Idiomas disponibles

Idiomas disponibles

Alarms and troubleshooting

Model
Alarm(s)
A980
Lisa 300/500
A981
Lisa 500 Fully
Automatic
A982 - 987
Lisa (VA131
2015, 2019)
Lexa (MN-111)
Lara (RIL-1)
Lina (RIK-1)
Lyla (RIN-210)
Cause(s)
Bad data connection.
Cycle started before the acoustic
tone.
Model selector in the wrong
position.
Bad serial connection.
Fault/leak in the pneumatic circuit.
Electronic boards and/or sensors.
Same as Lisa 500 Fully Automatic, but without the "A"
Turn off/on the calibration device.
Check the data cable connection.
Wait for the acoustic tone before starting the cycle.
Turn off/on the calibration device.
Check the model selector is in the correct position.
See also A982-987.
Turn off/on the calibration device.
Check the data cable connection.
Check the model selector is in the correct position.
Check the sterilizer circuit and the test connection for leaks. After fixing the
leak, run a vacuum test.
Pre-existing malfunctioning: check the "CYCLE HISTORY" menu of the sterilizer.
Action(s)
ENGLISH
DEUTSCH
FRANÇAIS
ESPAÑOL
20

Publicidad

loading