Descargar Imprimir esta página

Delete Calibration Data; Absorbance (Manual Range) - HACH LANGE 2100AN Manual Básico Del Usuario

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 101
Table 4 Diagnostic codes (continued)
Code
Display
23
Test results are shown.
24
Test results are shown.

Delete calibration data

To delete any calibration data entered by the user:
1. Turn off the instrument.
2. Push and hold CAL/Zero.
3. Turn on the instrument.
The CAL? light flashes. The instrument starts in Calibration mode.
4. Calibrate the instrument before use.
Flashing 9s
When manual ranging is selected, the display will flash all 9s when the
sample being measured is greater than the selected range.
When automatic ranging is selected, the display will flash 9s when the
sample is greater than the maximum range of the instrument. The
display will also flash 9s if Ratio is off and the measurement is greater
than 40 NTUs (268 nephelos or 9.8 EBCs). Turn Ratio on.
Flashing 0s
When manual ranging is selected, the display will flash all 0s when the
sample measured is less than the selected range.
When automatic ranging is selected, the display will flash all 0s when the
measurement is less than the range of the instrument or a negative
value. Calibrate the instrument.
• When measuring color, absorbance or transmittance, set the zero
reference point again.
• When measuring absorbance, make sure that the reading is positive.
To measure samples with negative absorbance, set the analytical
zero using the sample with the greatest absorbance and read the
All manuals and user guides at all-guides.com
Description
Keyboard test
Memory test
sample with the least absorbance. Record the reading as negative
absorbance.
English 21

Publicidad

Solución de problemas

loading