Intended use
This instrument is intended for use in laboratories, pharmacies, schools, businesses and light industry. It must only
be used for measuring the parameters described in these operating instructions. Any other type of use and
operation beyond the limits of technical specifications, without written consent from OHAUS, is considered as not
intended. This instrument complies with current industry standards and the recognized safety regulations; however,
it can constitute a hazard in use. If the instrument is not used according to these operating instructions, the
intended protection provided by the instrument may be impaired.
INSTALLATION
CAUTION: THE U-PALLET SCALE IS HEAVY, MOVING THE SCALE MAY
REQUIRE THE ASSISTANCE OF ANOTHER PERSON OR THE USE OF A
SUITABLE LIFTING DEVICE.
Unpacking
1. Remove the outer packaging materials.
2. Remove the U-Pallet scale from the packaging material.
3. Remove the 4 shipping feet (if installed).
4. Install the 4 load cell feet (if not already installed).
5. The i-DFP1500B1 and i-DFP1500C1 U-Pallet Scale may be moved by lifting at the rear handle and rolling
it on the two front wheels.
Positioning
Ensure the surface is flat in the area where the U-Pallet scale will be positioned, especially in the region of the load
cell feet. Small differences in height can be compensated by adjusting the load cell feet.
1. If an Indicator is included with the U-Pallet scale ensure that it is kept with the scale during positioning as
there may be an attached connecting cable.
2. Position the scale at the work site by lifting at the rear handle and rolling it on the two front wheels.
3. Check that the scale surface is level; check that the 4 load cell feet are in contact with the floor; make
adjustments by rotating the nuts on the load cell feet. Use an external level.
4. Ensure that the scale's output cable does not become trapped or crushed during lifting or positioning.
Connecting an Indicator
For users who buy the whole Defender 3000 floor scale:
The connection has already been made and no additional installation is required. If you want to connect the base to
another indicator in the future, you can refer to the wiring code in Table 1.
For users who buy the Defender 3000 base:
When connecting the base to an indicator that does not have the matching connector, install the optional load cell
cable adapter (P/N 30429931 sold separately) to the indicator using the wiring codes in Table 2. Then attach the
load cell cable to the cable adapter.
In addition, as an alternate wiring method, you can also cut off the connector, strip the wires and attach them to the
indicator using the wiring codes in Table 2.
The extension cable accessory (P/N: 30429931) cannot be used if the scale will be used in a legal for trade or
legally controlled application, because there is no provision to seal the connection between the indicator and
platform.
CAUTION: Ensure that the cable is protected against possible damage when routed to the Indicator.
Defender 3000 Series U-Pallet Scale