When using as motor alone mode, the torque bar will show on the
screen.
4.9 Battery Charging
There is a built-in rechargeable lithium battery in the motor han-
dpiece.Insert the power adapter plug into the Motor handpiece and
When the screen battery indicator fashes, it is charging. After char-
ging, please unplug the power adapter.
This device must use the original power adapter.
4.10 Replacing Battery
If the battery needs to be replaced,please contact local distributors.
Here is how to replace the battery.
a) Turn the motor handpiece power off.
b) Use tweezers etc. to open the rubber cover and then remove the
screw.
c) Gently separate the upper and lower cover of the Motor handpie-
ce.
d) Remove the old battery and disconnect the connector.
e) Connect the new battery and put it in the motor handpiece.
f) Fasten the upper and lower cover of the Motor handpiece, tighten
the screws and install the rubber cover.
4.11 Oiling of contra angle
Only the original oil injection nozzle can be used for oiling of contra
angle. The contra angle needs to be lubricated after cleaning and
disinfection, but before sterilization.
1. Firstly, screw the injecting nozzle into jet of oil bottle. (Around 1 to
3 circles)
2. Next, plug the nozzle into the end part of contra angle, and then
grease the contra angle for 2-3s till the oil fow out of contra angle
head part.
3. Vertically place the end part of contra angle more than 30 minutes
to let go the redundant oil under gravity.
Warnings
Cautions
a: To avoid the contra angle from fying out for the pressure use hand
to safely hold the contra angle while greasing.
b: Do not use a swirling nozzle. Swing nozzle can only be used for
injection of gas, not for oiling.
16