nents with a wet soft cloth for at least 3 minutes.
3. Wipe the surface of the component with a dry soft nap-free cloth.
Note:
a) The cleaning and disinfection must be performed within 10min
before use.
b) The disinfectant used must be used immediately, no foaming is
allowed.
c) In addition to 75% alcohol, you can use non-residue disinfec-
tants such as Oxytech from Germany, but you must respect the
manufacturer.
d) After cleaning and disinfecting the handpiece, you must install a
disposable isolation sleeve before use.
6.3.1.3 Post-Op processing
After each use, clean and disinfect the handpiece, charger, and base
Tools: Nap-free soft cloth, tray
1. Remove the contra-angle from the handpiece, place it in a clean
tray, and then remove the disposable isolation sleeve from the
handpiece.
2. Soak the nap-free soft cloth with distilled water or deionized
water, and then wipe all the surfaces of the components such as the
handpiece, charger, base, etc. until the surface of the component is
not stained.
3. Wet the dry soft cloth with 75% alcohol, and then wipe all surfaces
of the handpiece, charger, base and other components for 3
minutes.
4. Put the handpiece, charger, base and other components back into
the clean storage area.
Note:
a) The cleaning and disinfection must be performed within 10min
before use.
b) The disinfectant used must be used immediately, no foaming is
allowed.
c) In addition to 75% alcohol, you can use non-residue disinfec-
tants such as Oxytech from Germany, but you must respect the
manufacturer.
d) After cleaning and disinfecting the handpiece, you must install a
disposable isolation sleeve before use.