Verify performance of the AriaDx Real-Time PCR System
Prior to running your first assay on the Sistema de PCR en tiempo real AriaDx, Agilent recommends
running a system verification to verify the performance of the instrument and of all dyes you
intend to use in your assays.
To perform this system verification, run an experiment using a 96-well PCR plate (or set of PCR
tubes) that contains positive control reactions in one column and negative control reactions in
another column, across all rows of the plate that are in use in your assays. Use the same reagents
and dyes that will be used in your assays.
Agilent recommends performing this verification prior running your first assay and following any
intervention on the AriaDx instrument (such as, preventative maintenance operations, motor
calibration, and installation, replacement, or repositioning of an optical module).
Sistema de PCR en tiempo real AriaDx Guía de usuario y configuración
56