Fig. 15:
Filter change confirmation
Choose one of the two options
available on the Filter change con-
firmation screen:
1. If the filter was checked, and
replaced if necessary, press the
button beneath the
icon.
 The unit will begin the warm-
up cycle, the internal timer is
reset.
"
Resume normal operation.
2. To postpone checking the filter,
press the button beneath the
clock icon.
 The unit will complete the
warm-up cycle.
"
Resume normal operation.
Bosch Automotive Service Solutions GmbH
Maintenance | IR LD 1.0 | 31 | en
i
Postponing the filter check will
not reset the internal timer,
and the IR LD 1.0 will display
the Check sensor filter screen
each time the unit is powered
ON, until the filter is checked
and the first option (above) is
selected.
7.1.2
Refer to Figure 16 during the re-
placement procedure:
1. Unscrew the filter tip to access
the filter.
2. Remove filter from filter tip.
3. Ensure O-ring is properly
seated in the probe end.
4. Install new filter into filter tip.
5. Screw filter tip onto probe.
"
New filter is installed and ready
for use.
check
Fig. 16:
1 Probe end
2 O-ring
3 Filter
4 Filter tip
Filter replacement
3
4
Filter and probe components
SP01502142 | 14.03.2022
1
2