Lid Lock; Preselection; Preselection Of Speed / Rcf-Value - OHAUS Frontier FC5706 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

TM
Frontier
Centrifuge FC5706
M2
A-1
Figure. 18
During the run you can call up the rotor type at any time by pressing the key "Door Open" (7).

3.3.2 Lid lock

The lid should only be put down slightly. An electromagnetic lid lock closes the lid, at the same time the word
"open" (M2) disappears (refer to figure 18).
As a sign that the centrifuge is ready for starting, in the display "RPM | RCF" (A-1) the word "close" (M1) appears.
Simultaneously the word "rotor" (M3) is displayed, as well as the code number of the rotor, which is in the
centrifuge system, "no 71" (M4). With that, all rotor specific data, like e. g. max. speed, acceleration etc., are
adopted.
ATTENTION:
Before closing the lid please check if the rotor is tighten, and all 6 buckets have put in the swing out
rotor.
3.4

Preselection

3.4.1 Preselection of speed / RCF-value

This pre-selection is activated through the key "RPM | RCF" (4) (refer to figure 19 below). By pressing the key
once the word "rpm" (M5) flashes. By pressing the key twice the pre-selection of the centrifugal forces can be
selected. Then the flashing word "rcf" (M6) appears. You can set the desired values with the adjusting knob (1). In
the display (A-1) the regulated value is shown permanently, before, during and after the run.
A-1
4
5
Figure. 19
M3
M1
M5
M6
11
7
1
EN-7