Logic Max value / Logic Min. Value
If the hysteresis is not present, the upper and lower limits can be set; if the measurement
exceeds these values, an alarm is generated.
3.2.2
CALIBRATION
Figure 15 Calibration
This tab shows the calibration method for measurements; follow the instruction on the screen:
Insert probe in the buffer, wait for measure to be stable (in the circled areas), and then press OK. It opens
the follow window:
Figure 66 Set value
Set the buffer value and press OK.
Only the pH requires 2-point calibration (with 4 pH and 7 pH buffer solutions):
When you press "Calibrate", it opens the window of the other calibrations.