SYMBOLS
CE Medical Marking ref. Dir. 93/42
EEC and subsequent amendments
Class II device
Important: check the operating
instructions
"OFF" for part of equipment
"ON" for part of equipment
Complies with: European standard
EN 10993-1 "Biological Evaluation
of Medical Devices" and European
Directive 93/42/EEC "Medical
Devices". Phthalates free. In
conformity with Reg. (EC) no.
1907/2006
Minimum and maximum room
temperature
Minimum and maximum
atmospheric pressure
ELECTROMAGNETIC COMPATIBILITY
This device was designed to satisfy the currently required requisites for electromagnetic
compatibility (EN 60601-1-2:2015). Electro-medical devices require particular care
during installation and use relative to EMC requirements. Users are therefore requested
to install and/or use these devices following the manufacturer's specifications. There
is a risk of potential electromagnetic interference with other devices. RF mobile or
portable radio and telecommunications devices (mobile telephones or wireless
connections) can interfere with the functioning of electro-medical devices. For further
information visit our website www.flaemnuova.it. The Medical Device may be subject
to electromagnetic interference if other devices are used for specific diagnosis or
treatments. Flaem reserves the right to make technical and functional modifications
to the product with no prior warning.
Certification TÜV
Serial number of device
Manufacturer
Type BF applied part
Alternating current
Hazard: electrocution.
Consequence: Death.
Do not use device while taking a bath
or shower
Enclosure protection rating: IP21.
(Protected against solid bodies over
IP21
12 mm. Protected against access with
a finger;
Protected against vertically falling
water drops.)
Minimum and maximum air moisture
6