Operation
54
Transport of vacuum system
NOTICE
Status display (see 10/6.1 SC820G, SC840G [} 31])
à
Lights up pink if the vacuum system was switched on.
à
Illuminates red if there is a fault:
Signal duration
100% ON (continuous light)
100% ON (continuous light)
50% ON; 50% OFF
90% ON, 10% OFF
Tab.20: Fault signal via status display
For further information, see Chapter 12 Troubleshooting
[} 93].
Translation of Originial Operating Instruction, KNF 339416-339418 02/23
Vacuum system SC820G / SC840G
Damage may occur due to incorrect or
improper transport
If the collection flasks remain mounted
during transport of the vacuum system,
they can be damaged by careless han-
dling when the vacuum system is set
down.
à
Remove the collection flasks be-
fore transporting the vacuum sys-
tem.
à
If there are liquids in the collection
flasks, empty them or dispose of
the liquids in an environmentally
friendly manner.
à
Store the collection flasks in a safe
location.
à
Reassemble the collection flasks
after transport.
Fault type
Drive blocked
Emergency stop pressed
Temperature too high (drive)
Other fault