Technical Information
Vent Valve Operation – Automatic Opening Delay (AUTO TIME)
The vent valve is automatically opened in function of time
(WIN 126 = delay and WIN 147 = opening time).
After pump STOP command, the vent valve stays closed for the time set on WIN
126. After the delay, the valve opens for the time set on WIN147. If WIN147 = 0, the
valve remains open until next START command. If WIN 147 ≠0, the valve opens for
the time indicated. Then it closes until the next START command.
Reference to existing description: AUTOTIME, it's the default setting, WIN 125 = 2.