Technical Information
This submenu contains the settings related to the set point contact output with
relay.
TYPE: the output signal can be setted referring on one of following values:
FREQUENCY, default value
POWER drawn by the pump
TIME
NORMAL, when the pump has reaches Normal Operation mode
(Both HiFlow or HiComp)
SET VALUE, after selection of the analog output type, select the desired value from
0000 to 9999
HYSTERESYS, if a Hysteresys value is required, it can be adjusted between 00 and 99
MASK TIME Selectable between 00:00:00 and 99:99:99
OUTPUT LOGIC The contact logic is configured depending on Activation Type
parameter status.
If Activation Type is "high level active" the contact is configured as "normally
open" (it will close the circuit when activated);
if Activation Type is "low level active" the contact is configured as or "normally
closed" (it will open the circuit when activated).
196
Agilent TwisTorr Medium-TMP Rack Controller User Manual