All connections and disconnections must be performed by qualified and
appropriately trained staff.
4.1
Specifications and use
4.1.1 Nurse Call cable specification
Wires length = 5 cm Shrinkable tube = 2 cm
The unterminated side is composed of 3 wires, connected to Nurse Call signalling
system:
R: Stand-off contact
T: Working contact
C: Common contact
4.2
Connection to the holder
Connect the terminated Nurse Call cable to holder Nurse Call connector.
Pinch the cable into the slot provided for this purpose.
Connect the Nurse Call cable unterminated side to the hospital Nurse Call system
as described below.
Check that the Nurse Call system is functional by generating alarms, Refer to
Nurse Call quick check protocol
connected Nurse Call system.
To disconnect, unplug the Nurse Call cable.
4.2.1 Nurse Call system electrical schematics
Without pump alarms:
System operation
Pump is running without any alarm
Pump is OFF
No pump is installed on the holder
Pump is booting
10
. Ensure the pump alarm is transmitted on the
Contact
C-T
Electrical continuity
C-COM-NO-T