Descargar Imprimir esta página

Fresenius Kabi COM Instrucciones De Uso página 17

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 124
time for each step is 3 minutes for bactericide activity (surface remain visibly wet
for 3 minutes). Respect the indicated contact time from the manufacturer
recommendations for the required antimicrobial activity.
5.
Wipe down the cables.
10
Transport, storage and recycling conditions
10.1 Storage and transport conditions
Before transport, disconnect and remove the power cable from the holder.
Disconnect all other cables.
Check power cable and Nurse Call cable are operational after transport.
The holder should be used under specified storage and transport conditions (see
environment
) to ensure holder performance and to avoid holder malfunction.
10.2 Storage
Please make sure the holder is stored in an appropriate manner so as to avoid holder
malfunctioning.
INFORMATION
The storage area must be clean, organized and compliant with the
storing conditions mentioned above.
The holder must be handled with care during storage.
CAUTION
The holder must be cleaned and disinfected prior to storage (see
and disinfecting
10.2.1 Prepare the device for storage
In order to prepare the device before storage, proceed as specified below:
1.
Disconnect power cable and all other cables.
2.
Remove the pump and holder from pole or rails.
3.
Clean the holder (see
4.
Handle the holder with care and store it in a compliant area.
10.2.2 Install the device after storage
INFORMATION
We recommend that a quick check protocol is performed (refer to per
pump IFU, section Quick Check Protocol) when the device is installed
after transport, in case of prolonged storage, or before being used on a
new patient.
To ensure Nurse Call is functional, we recommended that the
quick check protocol
before use on a new patient, after cleaning or after moving) and shall
be performed by qualified and appropriately trained staff.
).
Cleaning and disinfecting
is applied regularly after installation (especially
).
Use
Cleaning
Nurse Call
17

Publicidad

loading