2.6 Draft Shield Installation (Carat models only)
Lay the dome shaped draft shield over the notched metal ring with the open end facing
towards the back of the balance. With downward pressure rotate the draft shield
allowing the bottom edges to slide under the notches of the metal ring. Continue
rotating the draft shield until the opening is facing forward. See Figure 3-1.
Figure 3-1. Draft Shield Installation.
2.7 Initial Calibration
Power on the balance by pressing On-Zero. Allow at least one minute for the balance to
stabilize before calibrating. Press and hold the Print/Cal button until [CAL] is
displayed. The display flashes [-C-] while zero reading is taken. The display flashes
the required span mass. Put the indicated calibration mass on the pan and press Yes.
The display flashes [-C-]. When calibration is complete, [DonE] is displayed.
Note: Calibration is also available in the Setup menu, see section 4.2.1.