overcharging so the balance can remain connected to the AC power. When the battery is fully
charged the battery indicator on the display will stop blinking.
To remove the rechargeable battery option and install C cell batteries, reference the Recharging
Battery Option instruction manual for step by step instructions as well as disposal instructions.
CAUTION: Risk of explosion can occur if the rechargeable battery is replaced with
the wrong type or if it is not properly connected.
Note:
After power on, it is recommended to warm up the balance for at least 5 minutes before using it.

2.4 Initial Calibration

When the balance is first installed it should be calibrated to ensure accurate results.
Press and hold Menu-Cal until [mMeNU] (Menu) is displayed. When the button is released, the display
will display [.C.A.L.]. Press Yes to accept, [SpaN] will then be shown. Press Yes again to begin the
span calibration. [--C--] blinks while zero reading is stored. Next, the display shows the calibration
weight value. Place the specified calibration mass on the pan. [--C--] blinks while the reading is
stored. The balance returns to the previous application mode and is ready for use.
Required Span Calibration Mass (sold separately)
Capacity Mass
610 g 300 g
2100 g 1000 g
NVE