Electrical connection
Two fittings are provided on the left side of the enclosure: the AC Power bulkhead
fitting for power cable, and the M800 patch cable connection. Feed the power cable
through the bulkhead fitting on the side-wall. AC power connections are positioned
on the right side of the printed circuit board. See Figure 3 for terminal wiring. Ensure
line fuse is properly installed. See product label for specified fuse size.
Figure 3
Introduce Sample Flow
Slowly open the sample point isolation valve (not supplied), to start flow to the
sensor. Allow approximately 3-5 minutes for the sensor to fill, and observe flow at
the drain tube. Once there is flow, check that there are no leaks inside the sensor
enclosure. Allow the sensor to initially flush with sample water (4-24 hours is recom-
mended). Close front cover and connect the M800 patch cable to the connector on
the bottom left side of enclosure.
Figure 4
HIGH CAPACITY INLET FILTER
ASSEMBLY & DRAIN