5000TOCi Sensor Installation and Start-Up
Installation
Mount the 5000TOCi in an appropriate location. a minimum clearance dimension
of 18" (45.7 cm) is required on the left side of the sensor for Uv lamp change-out.
Two plumbing connections are provided on the right side of the enclosure. The lower
connection is labeled 'Sample Inlet' and the upper connection 'Sample Outlet'. a third
connection, a safety drain, is located on the bottom of the enclosure. Figure 2 shows
a typical installed configuration. The installation kit provided includes installation
tube assemblies, a sample conditioning coil and a high capacity inlet filter. Attach
the high capacity inlet filter to the sample inlet connection as shown in Figure 4.
Figure 2
Connect the 0.125" (3 mm) PTFE tubing to a sample point isolation valve (not
supplied). Warning: For inlet pressure above 85 psi (5.9 bar), a pressure regulator
(Thornton p/n 58 091 552) is required. Flush sample tubing before installing at
the sensor. Close the isolation valve. Attach the other end of the PTFE tubing to the
sample conditioning coil then attach the sample conditioning coil to the high capacity
inlet filter.
Connect the stainless steel drain tube (Thornton p/n 58 091 553) as shown in Figure
2 and direct outlet to an atmospheric drain.