To enter the technical menu switch On the tool and after the page with the TECNO-GAZ
logo is displayed wait for the next page indicating that the tool is operating.
Hold down the STOP button and at the same time press the OK button twice to enter the
technical menu.
OK
OK
OK
OK
TECHNICAL MENU
TUB : water level into the tub
DETERGENT: detergent tank level
PT100°C : temperature measured by the probe on the tank
From peripherals section press OK to scroll down options
Press SELECT to select activation/deactivation of components
PUMP
CHARGE SV
DRAIN PUMP
ULTRASONIC (warning: never start ultrasounds with empty tub)
RS TUB (warning: never start the resistance with empty tub)
RS DRYING (warning: never start the resistance without starting the
FAN first)
FAN
EXIT
Quick test cycle
This section shows the total cycles performed by the tool
The tool goes back to the operating mode.
DEVICES
TEST CYCLE
COUNTER
EXIT
ENGLISH