ENGLISH
ERR 010
Check the continuity in wiring and the operation of the level sensor
Check that there is no leak in the pipes inside the machine
Check that there is no vacuum in the drain tube
Replace the Main Board
ERR 011
Check that the water charge tap of your water system is open
Check the system pressure (1.5 bar minimum - 2 bar maximum)
Check that there are no obstructions in the inlet water pipe
Check that there is no water in the tank and then start a new cycle
Check the cleanliness and integrity of the level probe
Clean vessel filter, water inlet filter on EV charge
Check operation EV charge (enter the technical menu and enable the functioning of the
single device) if necessary replaced EV charge
Check continuity in wiring that connects EV charge with Main Board and if necessary
replace it
Check the WATERBLOCK (OPTIONAL) is not blocked
Replace the board
30
During tools washing the level of probe doesn't marks MAX
It occurs if there are the following conditions:
• If it takes more than 4 minutes (second water inflow)
HOW TO OPERATE
READING ERROR OF WATER LEVEL
HOW TO OPERATE
PROBLEM DURING WATER INFLOW
PHASE