ENGLISH
RESET
In case of error reset the machine, hold down
START
Start the pump for 30 second, hold down
PUMP
ERR 005
Check that the water charge tap of your water system is open
Check the system pressure (1.5 bar minimum - 2 bar maximum)
Check that there are no obstructions in the inlet water pipe
Check that there is no water in the tank and then start a new cycle
Check the cleanliness and integrity of the level probe
Clean vessel filter, water inlet filter on EV charge
Check operation EV charge (enter the technical menu and enable the functioning of the
single device) if necessary replaced EV charge
Check continuity in wiring that connects EV charge with Main Board and if necessary
replace it
Check the WATERBLOCK (OPTIONAL) is not blocked
Replace the board (L)
ERR 007
It occurs if the PT100 probe does not perceive an increase of 1°C in 5 minutes
Replace the PT100 (K)
Replace the board (L).
Check the resistance of the silicon heaters (I) (correct medium value 150 Ω)
28
PROBLEM DURING WATER LOADING
It occurs if there are the following conditions:
If it takes more than 4 minutes
HOW TO OPERATE
PROBLEM DURING WATER HEATING
HOW TO OPERATE
for 8 seconds.
for 8 seconds.
PHASE
PHASE