4.2
Assembly – guide light
⇒ Disconnect the power plug.
To mount the guide light you will need the trimmer Allen key.
This is located on the left side of the unit behind the door.
⇒ Open the door locks.
⇒ Remove the trimmer door and place it to one side.
⇒ Remove the Allen key from the holder.
⇒ Lift off the cover strip with a small screwdriver. There is a small
slot on the back of the cover strip for this purpose.
⇒ Use pointed pliers or small diagonal pliers to break open the
plastic at the intended location on the cover strip.
⇒ Insert the connection cable (5, Fig. 1) in the connection point on
the trimmer.
If necessary, use a small screwdriver to press the connector
inwards until it clicks into place.
⇒ Attach the cover strip again. In doing so, guide the cable through
the opening created previously.
⇒ Attach the guide light. The ridges on the front and rear must
engage in the slot on the unit.
The cable should not be pinched!
⇒ Tighten the clamping screw (6, Fig. 1) on the back of the guide
light only slightly, as the guide light still has to be aligned.
⇒ Connect the power plug.
⇒ Switch on the guide light at the sensor button (4, Fig. 1).
⇒ Align the guide light by moving it to the right/left and turning the
black adjustment wheel so that the laser beam is aligned with
the median palatine raphe of the orthodontic model table.
⇒ Tighten the Allen screw finger-tight.
Do not overtighten the Allen screw!
- 3 -
en