RAM® Dual Cable Series
RAM Dual Cable LNC
Reusable
INDICATIONS
The RAM® Dual Cable LNC cables have the same indications for use as the applicable sensors. Please see the compatible
sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
The RAM Dual Cable LNC cables are for use with the rainbow Acoustic® Respiration Sensors )RAS( and LNCS® sensors.
Refer to Fig. 1. The RAM Dual Cable LNC provides two )2( types of sensor connections:
1. )A( A connection to the Acoustic Respiration Sensor )RAS( and
2. )B( A connection to the LNCS sensor.
The RAM Dual Cable series have been verified using Masimo® rainbow SET® Technology.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module Operator's Manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable
and discard if cracks or signs of deterioration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable
and sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cables to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or
disconnecting either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize
the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may
damage the electrical components, potentially leading to patient harm.
• Caution: Replace the sensor when a replace sensor message is displayed, or when a low SIQ message is
consistently displayed while monitoring consecutive patients after completing the low SIQ troubleshooting steps
identified in the monitoring device operator's manual.
• Note: The cable is provided with X-Cal® technology to minimize the risk of inaccurate readings and unanticipated loss
of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time. Replace the cable when the
patient monitoring time is exhausted. Some legacy monitoring devices do not distinguish between the cable and the
sensor. In these devices, a replace sensor or similar message could pertain either to the sensor or the cable.
INSTRUCTIONS
NOTE: Connect the RAM Dual Cable LNC to the rainbow Acoustic Monitoring® enabled device before connecting
the sensors.
A. Connecting to the rainbow® Acoustic Monitoring Enabled Device
1. Connect the device connector of the patient cable to the mating connector of the device.
2. Ensure the patient cable connector is fully engaged with the device. A tactile or audible click of connection should be heard.
B. Connecting the RAM Dual Cable LNC to the Sensors
Connect to a RAS sensor:
1. Refer to Fig. 2. Orient the arrow on the RAM Dual Cable acoustic connector )A( with the arrow on the RAS sensor )B(.
2. Insert the sensor connector completely into the acoustic connector until it locks in place.
Connect to a LNCS sensor:
1. Refer to Fig. 3. Open the LNCS sensor's protective cover.
2. Insert the RAM Dual Cable LNC connector completely into the sensor connector )A(.
3. Completely close the protective cover )B(.
C. Disconnecting from the Sensors
NOTE: Pull on the connectors only, do not pull the cables.
Disconnect from a RAS sensor:
1. Refer to Fig. 4. Hold the center of the RAM Dual Cable LNC's acoustic connector )A( and the RAS connector )B(.
2. Pull the connectors apart to disconnect.
Disconnect from a LNCS sensor:
1. Refer to Fig. 5. Lift the LNCS sensor's protective cover to gain access to the sensor connector )A(.
2. Pull the connectors apart to disconnect )B(.
NOTE: The RAM Dual Cable LNC can be connected/disconnected while the device is operational.
DIRECTIONS FOR USE
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
3
en
Non-Sterile
7490F-eIFU-0522