8. Commissioning the unit
8.1 Before starting work
Before starting surgery, the HBS 300i x-touch electrosurgery unit is connected via the supply cable to
the mains socket with PE conductor. The current cable may only be attached or removed when the
device is switched off. The connecting socket for the power cable is located at the back of the housing.
Foot switches are connected at „mono1", „mono2", „bi1" or „bi2" . The terminals are located at the back
of the housing.
Handles and electrodes are connected correspondingly at the front of the housing to "MONOPOLAR1",
"MONOPOLAR2", "BIPOLAR1" or "BIPOLAR2".
8.2 Switching on
The system is switched on by actuating the main switch (ON/OFF). The device system automatically
controls all functions of the device after switching on and also during the entire working time. However,
high-frequency current only flows once a button is activated at the handle or foot switch.
Consequently, accessories can also be connected to the unit when it is switched on.
However, extreme care should always be taken when doing so!
Always ensure that the system is not inadvertently activated during
installation by pressing the foot switch or the button in the handle.
44
Before first using the electrosurgical unit, we recommend
familiarizing employees with the operating characteristics of the
different operating modes and power settings by carrying out
practice exercises using fresh meat (beef, pork).
When using an electrosurgical unit which produces HF current, always
be aware of two fundamental principles:
◼
Current flows through everything located between the active
and the neutral electrode (monopolar)