Table 6-2
Performance verification tests (continued)
Step
Test function
11
Press
to select frequency counter
12
Press
to select divide-by-100 frequency
counter mode
13
Turn the rotary switch to the
14
Press
to select temperature measurement
[h][m]
15
Turn the rotary switch to the
16
Press
to select ACA measurement
17
Turn the rotary switch to the
18
Press
to select AC mA measurement
Keysight U1253B User's and Service Guide
Range
999.99 kHz
[f]
99.999 MHz
10.000 nF
[g]
position
100.00 nF
1000.0 nF
10.000 F
100.00 F
1000.0 F
10.000 mF
100.00 mF
–40 °C to
1372 °C
500 A
position
5000 A
500 A
[a]
5000 A
position 50 mA
440 mA
50 mA
[a]
440 mA
Performance Tests and Calibration
Reference signals/
values
33250A output
200 mVrms,
100 kHz
600 mVrms,
10 MHz
5520A output
10.000 nF
100.00 nF
1000.0 nF
10.000 F
100.00 F
1000.0 F
10.000 mF
100.00 mF
0 °C
100 °C
500 A
5000 A
500 A, 1 kHz
5000 A, 1 kHz
50 mA
400 mA
50 mA, 1 kHz
400 mA, 45 Hz
400 mA, 1 kHz
6
Error limits
± 52 Hz
± 5.2 kHz
± 108 pF
± 1.05 nF
± 10.5 nF
± 105 nF
± 1.05 F
± 10.5 F
± 105 F
± 3.1 mF
± 1°C
± 2°C
[i]
± 0.3 A
[i]
± 3 A
± 3.7 A
± 37 A
[i]
± 80 A
[i]
± 0.65 mA
± 0.37 mA
± 3 mA
± 3 mA
177