Table 6-4
List of calibration items (continued)
Function
DC mA/DC A
AC mA/AC A
Keysight U1253B User's and Service Guide
Range
Calibration item
Open for all ranges
OPEN
50 mA
30.000 mA
500 mA
300.00 mA
Move the positive test lead from the A.mA terminal to the A terminal.
Caution: Connect the calibrator to the multimeters A and COM terminals before applying
3 A and 10 A.
5 A
3.0000 A
10 A
10.000 A
(done)
50 mA
3.000 mA (1 kHz)
30.000 mA (1 kHz)
500 mA
30.00 mA (1 kHz)
300.00 mA (1 kHz)
Move the positive test lead from the A.mA terminal to the A terminal.
Caution: Connect the calibrator to the multimeters A and COM terminals before applying
3 A and 10 A.
5 A
0.3000 A (1 kHz)
3.0000 A (1 kHz)
10 A
3.000 A (1 kHz)
10.000 A (1 kHz)
(done)
Performance Tests and Calibration
[a]
Reference input
Unplug all test leads or shorting plug,
and leave the terminals open
30 mA
300 mA
3 A
10 A
3 mA, 1 kHz
30 mA, 1 kHz
30 mA, 1 kHz
300 mA, 1 kHz
0.3 A, 1 kHz
3 A, 1 kHz
3 A, 1 kHz
10 A, 1 kHz