8) CLEANING INSTRUCTIONS
Do not use decontamination or cleaning agents which could cause a hazard as a result of a reaction with
parts of the material contained in it.
Clean oven after each use. Any gutta-percha remaining on the holder or in the heating chamber may be
effectively removed with Orange Solvent (generic dental solvent) on a cotton pledget. Press and hold
both buttons at once for 2 seconds to put the oven into the cleaning mode. This will cause both heaters
to heat up to a medium temperature of 140°F/60°C. The indicator lights around the buttons will show that
the heaters are on (constantly on).
A signal "beep" and flashing indicator lights around the buttons will warn that the desired temperature has
been reached. Disconnect the mains power plug and clean the heater chambers with a wire brush from
bottom side as indicated.
To abort the cleaning mode without disconnecting the mains power plug and to enter the normal mode
again press either button for 2 seconds.
If there is any doubt about the compatibility of decontamination or cleaning agents, please contact the
cleaning agent's manufacturer or the distributor.
9) TECHNICAL SPECIFICATIONS
Voltage rating
Standby power
Normal use power
Fuse rating
Equipment mass
Equipment dimensions
The equipment is designed to be safe for indoor use under the following environmental conditions:
Operating temperature
Max. operating humidity
Storage temperature
Storage humidity
Max. altitude
Transient overvoltages
Pollution degree
Mains voltage fluctuations
Electromagnetic environment
Thermaprep®
Obturator Oven
115 or 230 VAC (± 10%), 50/60 Hz
<1.5 W
380 W max.
4AT 250 V (non-replaceable)
0.56 kg
155 x 178 x 81 mm
+5°C to 40°C, +41°F to 104°F
80% at +31°C to 50% at 40°C (lin.), 80% at +87°F to 50% at 104°F (lin.)
-20°C to +65°C, -4°F to +149°F
20 to 80%, no condensation
2000 m
Category II
2
Up to ± 10% of the nominal voltage
Basic
END-IFU-THERMAPREP-V01-WEB-EUR-multilingual-2021-05