Startup
6.1
Device setup
• Installations-Layout
• Place the Vacuson 40/60 suction pump and all required accessories and instruments on an even, non-slip
surface and make sure you have good access to all controls.
• The installation of the device in close proximity to other devices is prohibited due to EMC – please see sec-
tion 3.1 and the manufacturer's EMC declaratiion in the appendix of this manual.
• Do not allow the operating range of the device (including cable) and the connected instrument to be com-
promised by limiting factors.
• The manometer must be fully visible at all times.
• The foot switch with pedal must be placed within stepping distance between the patient and the surgeon.
• It must be explicitly ensured that no objects can fall on the pedal.
• The power plug at the rear of the device must be accessible at all times.
• The ventilation slots at the housings bottom of the Vacuson 60 LP must be kept clear in order to prevent
temperature from becoming excessive.
• While in operation the device must be at least 1 meter above ground.
6.2
Connection to the power supply
Before switching on, make sure that the power supply unit of the device matches the country's specific
service voltage!
The power supply unit of the Vacuson 60 LP pump is not swichable to the country specific service voltage.
The device has to be ordered according to the country specific service voltage.
In order to prevent the risk of an electric shock, the device may only be connected to a power network
with a PE protective ground conductor.
Use only a certified power cord to connect the device to the power supply.
The power plug socket is located at the rear of the device.
8
Operating instructions Vacuson 60 LP, REF 31793, V34/17
Vacuson 60 LP