Descargar Imprimir esta página

Cuda surgical Elite 1000 Manual De Usuario página 10

Ocultar thumbs Ver también para Elite 1000:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 49
The hour meter located on the rear of the unit shows the total elapsed running time of the unit.
When the lamp is replaced, reset the hour meter to track total lamp usage hours on the
replacement lamp.
7. CLEANING AND DISINFECTION
WARNING: Ensure that the illuminator is de-energized and disconnected from mains power before
attempting to clean and disinfect.
The illuminator can be wiped down with commercially available cleansers commonly used for
disinfection of electronic equipment in hospitals such as ethyl or isopropyl alcohols, disinfecting
sprays containing quaternary ammonium compounds, or hydrogen peroxide.
WARNING: DO NOT use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid, or similar products. DO NOT use acetone, methyl ethyl ketone, or
halogenated / chlorinated hydrocarbon solvents or cleansers containing any of these restricted
compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids onto the device. Do not
allow liquids to enter the device seams or ventilation openings.
Follow all applicable bloodborne pathogen procedures as required by OSHA and/or your
hospital when cleaning and disinfecting the product.
WARNING: The illuminator is not sterilizable. DO NOT attempt to autoclave the device.
8. MAINTENANCE, SERVICING, REPAIR & WARRANTY
Performance of preventative maintenance is not essential. Defective items or equipment are to be
serviced and repaired exclusively by persons authorized by the manufacturer. All repair work shall
employ original manufacturer's parts only.
8.1 Lamp Replacement
WARNING: Ensure that the illuminator is unplugged from mains power and de-energized before
attempting to replace the lamp.
WARNING: Ceramic Xenon lamps are at high internal pressure when cold and at operating
temperature. Ceramic Xenon lamps may unexpectedly rupture resulting in discharge of hot fragments
of quartz and/or glass and metal. Only handle lamps with protective covers in place.
WARNING: Do not change lamp while touching patient.
NOTE: This illuminator is designed for exclusive use with the ST Technologies
lamp module. Failure to use this lamp could void warranty.
Turn unit off and unplug from mains power. Open lamp door located on side of illuminator.
Move the lamp positioning lever (found to the right of the lamp) clockwise, from horizontal to
vertical, and pull the lamp cartridge out.
Grasp only the horizontal top and bottom grip flanges of the plastic housing of the lamp, pull
straight out to remove the lamp. Insert a new one (lamp pointed to the left), pushing firmly to
assure full engagement to the power connectors. Lower the lamp engagement lever
counterclockwise, from vertical to horizontal, and observe that the lamp rises about 3mm as it
moves to its proper operating position.
Log the serial number and hours on the hour meter. Close the lamp compartment door. Re-
connect the power cord and turn the light source on according to 6.1. Reset hour meter to zero by
depressing hour meter reset button on rear panel
®
LIT-217 CUDA
Surgical
Rev.F
Page 10 of 98
®
P/N: SYC0300
(English)

Publicidad

loading