Mensaje de error
Superior: pantalla del panel
Código
de control
de error
Inferior: pantallas de la
estación de trabajo y del
controlador
ERROR
NO FRONT PANEL
Autosampler: The front panel
0x5226
is not installed properly or the
sensor for the front panel has
failed. Install the front panel
properly.
ERROR
ADJUST PLATE
Autosampler: Analysis was
0x5227
stopped because the position
of needle to plate was not
adjusted. Adjust the position
of the needle.
ERROR
ADJUST INJ.PORT
Autosampler: Analysis was
0x5228
stopped because the position
of needle to injection port was
not adjusted. Adjust the
position of the needle.
WARNING
RINSE SOLUTION
0x5229
Autosampler: Rinse solution is
low. Add rinse solution.
WARNING
RINSE SOLUTION R0
0x522A
Autosampler: Rinse solution
R0 is low. Add rinse solution.
WARNING
RINSE SOLUTION R1
0x522B
Autosampler: Rinse solution
R1 is low. Add rinse solution.
Causa principal
Asegúrese de que el panel frontal
esté correctamente colocado.
r
No se ha podido
detectar el panel
Si el error persiste, pulse el [Power
frontal.
button] del instrumento para
APAGAR la fuente de alimentación y
póngase en contacto con un
representante de ventas o servicio
de Shimadzu.
La posición de la aguja
Ajuste la posición de la aguja según
no se ha ajustado en
la placa.
función de la placa.
La posición de la aguja
no se ha ajustado en
Ajuste la posición de la aguja según
función del puerto de
el puerto de inyección.
inyección.
El nivel de solución de
Añada una solución de aclarado
aclarado restante es
adicional.
bajo.
El nivel de solución de
Añada una solución de aclarado R0
aclarado R0 restante es
adicional.
bajo.
El nivel de solución de
Añada una solución de aclarado R1
aclarado R1 restante es
adicional.
bajo.
5.4 Lista de mensajes de error
Contramedida
"Instalación y extracción
del panel frontal" del
manual de instrucciones
del inyector automático.
211
Serie CL de Nexera
5