In the unlikely event that something should go wrong, we have incorporated a number of safety features to ensure that your autoclave remains
safe at all times.
Safety features.
1. Located to the rear of the lid, beneath the cover, is a spring called
the G asket O ffset Device ( G O D Spring), designed to prevent
pressure building up if the lid has been incorrectly fitted.
D O N OT TAMPER WITH THIS SAFETY DEVICE
2. If for any reason, the temperature falls below the minimum
required sterilizing temperature, resulting in the Sterilizing
Temperature light (L3) switching off, the cycle timer will re-start
from zero once the correct temperature has been restored.
3. If there is an electrical or electronic failure resulting in a build up
of pressure - in excess of normal operating pressure - one or all of
the following safety features will be activated.
i)
Depressurization Valve (A) will loudly and rapidly "vent"
steam.
ii)
The gasket will "extrude" through the slot in the rear of
the lid rapidly releasing excess pressure and steam.
iii) A non - resettable thermal fuse located in the base of the unit
will "melt' at a pre-determined temperature, disconnecting
the power.
Technical Specifications - Standard Body Autoclaves
Height...............................335 mm
Width................................340 mm
Net Weight........................4.5 kilos
Height...............................420 mm
Width................................340 mm
Net Weight........................4.5 kilos
Height...............................582 mm
Width................................340 mm
Net Weight........................4.5 kilos
Fuses - Located under the control module, fuses
F10A, 32 x 6.3mm, ceramic sand filled, Mains plug top fuse (User
replaceable), F13A to BS1362 UK O NLY.
Rating - Models are rated continuously for intermittent use.
Body
- Deep drawn aluminium.
Lid - Aluminium.
Heater
- Externally surface mounted mechanically fixed electric
element.
Temperature Cut O ut - Thermal fuse.
Pressure - C alibrated pressure release valve.
Tech n ic a l Sp ecific a tio ns
C apacity............................9 litres
Internal chamber
Dimensions (d/h).............210 / 230 mm
- Extended Body Autoclaves
C apacity...........................12 litres
Internal chamber
Dimensions (d/h).............210 / 270 mm
- Super Extended Body Autoclaves
C apacity............................16.8 litres
Internal chamber
Dimensions (d/h).............210 / 444 mm
En g lish
Should any of the devices listed activate, please observe the
following steps:
a)
Do not touch the unit.
b)
Switch off at the wall socket and un-plug
c)
Allow temperature and pressure to drop
before
i) touching the unit
ii) removing your instruments
d)
Do not attempt to re-start the unit
e)
Arrange for an immediate service.
Max. Single Fault Temperature - 133.3°C
O ver Voltage C ategory - G roup ll
Pollution Degree - G roup 2
Environment C onditions - indoor use - temperature 5°C to 40°C -
altitude up to 2000m - maximum relative humidity 80% for
temperatures up to 31°C decreasing linearly to 50% relative
humidity at 40°C . - mains supply voltage fluctuations not to exceed
+ 10% of the nominal voltage
Input C onnections - Mains inlet socket 'hot' format conforming to
IEC 302.
Safety Shut Down - See 'Temperature Cut O ut'.
Packaging - All packaging materials are recyclable.
8
Max' Instrument
length............................... 228 mm
Max' Load Weight...........
Max' Instrument
length...............................290 mm
Max' Load Weight...........
Max' Instrument
length...............................460 mm
Max' Load Weight...........4.0 kilos
3.0 kilos
4.0 kilos