The "hold" mode is used to temporarily stop the flow of feed:
• to change the feeding program (installed parameters),
• to change the feed container,
• to administer medication without switching the pump off,
• to silence an alarm and correct problems.
After 3 minutes a two tone audible alarm sounds and the message "PUSH STRT" appears in the display. Press "start/stop" to
stop the alarm and to extend the hold mode by a further 3 minutes or press "start/stop" twice (3 beeps will be heard) to resume
programmed settings.
r
fill set
e
The FILL SET function should only be used to fill an empty (air-filled) feeding set. Use of the FILL SET function at any other time
may cause the pump to incorrectly calibrate.
a
Ensure the feeding set is not connected to the patients feeding tube when the pump is in the hold mode:
Press and hold the "fill set" key for 2 seconds to activate the "fill set" manoeuver. Release the button as soon as the pump gene-
rates a beep and starts pumping at a flow rate of approx. 700ml/h. During this manoeuver the message "FILL SET" appears in
the display. The pump will automatically stop when the Flocare
this "fill set" manoeuver can at anytime be stopped by pushing the "fill set" key a second time. The pump will return to the hold
mode when the "fill set" manoeuver is complete or stopped.
The pump has a liquid crystal display (L.C.D.) with large alphanumeric characters, smaller symbols, words and a back light.
The following information can be found in the display:
402981282_r1.indd 91
402981282_r1.indd 91
The pump beeps and carries out a short self test, showing the pump serial number in
8 digits. Verify that all display segments and symbols are active as shown in figure 6
(see next page).
The pump displays the total volume delivered since the memory was cleared and swit-
ches in the hold mode ready for programming.
SWITCHING "OFF"
Press the
key and keep it pressed for 2 seconds. A continuous alarm will be heard
and the pump switches off.
The feeding program (installed parameters) and total volume administered since the last clearance
will be retained in the pump's memory.
If the pump shuts down due to low battery voltage, the memory will be retained for 24 hours.
"HOLD" mode
To
temporarily
pause
operating, press the "start/stop" key once. Three beeps are heard and the run symbol disappears.
The programmed flow rate (ml/h), volume (volume ml), the bolus volume (bolus ml), the bolus interval
(int h) and the administered volume (info) are retained.
FILLING THE PUMP SET
The FILL SET function offered by the Infinity pump is used to fill a (new) feeding set with nutrition
(or water as the case may be). When the FILL SET function runs all alarms are deactivated helping
avoid alarms being inadvertently triggered (e.g. air alarm).
DISPLAY
the
pump,
or
switch
Pack Infinity
pump set is completely filled with feed. However
®
-91-
into
"hold"
mode
whilst