Descargar Imprimir esta página

Nutricia Flocare Infinity+ Instrucciones De Uso página 94

Ocultar thumbs Ver también para Infinity+:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 37
• Flow rate (ml/h), volume (ml) and interval (h) are displayed through the large characters. Words below describe what the number
relates to (ml/h, total volume, volume bolus, interval bolus, total volume del). The pump also displays messages, for example
"END OF DOSE" will appear when the pump has finished delivering a single feed dose.
Figure 6
The back light of the display turns off 10 seconds after the last key is pressed.
The back light will switch on for 10 seconds when the charger is connected to the mains.
• The Flocare
Infinity
+ pump can be used for continuous or dose feeding.
®
• Insert the pump set in the pump, switch the pump on and fill the pump set when necessary (see paragraph: filling the pump
set).
• The display shows the total volume delivered since the pump was last cleared.
• If necessary clear the volume delivered by pressing the "clr" key.
• The last installed flow rate (ml/h) is now displayed.
• The pump is now in the "hold" mode and is ready for programming.
• We strongly recommend that for those patients where interruptions or delay in therapy could affect their health status, the
DOSE setting of the Infinity pump must be used and programmed.
ml/h
• Adjust the flow rate (ml/h) if required by using the "+" or "-" key. Hold either key down to change rapidly.
• If another parameter is indicated in the LCD, it will be necessary to first press the "ml/h" key followed by the "+" or "-" key to set
the flow rate.
• The flow rate ranges from 1 - 400 ml/h in 1ml increments.
• The flow rate slows down and stops shortly at 50 - 125 - 250 ml/h.
• Press the "volume ml" key and make sure the value is disabled by pressing the "clr" key. The display will indicate the word "CONT".
• Press the "bolus ml/int h" key and make sure the volume bolus value is disabled by pressing the "clr" key. The display will indicate
the word "NO". This means the pump will run in a continuous mode.
• Start the pump by pressing "start/stop".
• The arcs around the word "run" start rotating.
• The flow rate (ml/h) is displayed whilst functioning.
402981282_r1.indd 92
402981282_r1.indd 92
• The arcs around the run symbol rotate when the pump is running.
• The remaining battery capacity (see paragraph: battery operation).
• The wall plug symbol indicates that the charger is plugged in.
PROGRAMMING THE PUMP
CONTINUOUS FEEDING
+
start/stop
1 ml/h
-92-
EN
06/01/23 4:50 AM
06/01/23 4:50 AM

Publicidad

loading

Productos relacionados para Nutricia Flocare Infinity+