Problem
LCD: no indication.
Not enough
measurement precision.
Reduced or zero flow rate.
The meter does not count,
but the flow rate is correct.
SAMOA Industrial, S.A. · Pol. Ind. Porceyo, I-14 · Camino del Fontán, 831 · 33392 - Gijón - Spain · Tel.: +34 985 381 488 · www.samoaindustrial.com
Dimensions
Possible cause
Bad battery contact.
Wrong K FACTOR.
The meter works below.
minimum acceptable flow rate.
TURBINE blocked.
Incorrect installation of
gears after cleaning.
Possible electronic card
problems.
Remedial Action
Check battery contacts.
With reference to paragraph H,
check the K FACTOR.
Increase the flow rate until an
acceptable flow rate range has
been achieved.
Clean the TURBINE.
Repeat the reassembly
procedure.
Contact your dealer.