PeratInG the UnIt
H. Pause / standby function
• The (B) button offers a "pause / standby function" for the
entire operation of the device. Both the handpiece rotation
and the spray function are interrupted. The values set remain
unchanged during the pause period. During the standby mode,
the icon
the display is illuminated.
• By pressing this button again, the device continues to run with
the previous settings.
DE
• The same function can be achieved by actuating the rocker
switch (V) at the end of the handpiece (U).
EN
I. Saving speed values / memory function
A selected motor speed, spray intensity or direction of rotation
FR
can be saved by pressing and holding any memory button
(K) for at least 3 seconds. The following memory values have
IT
already been factory-set:
M1 = (8.000 rpm, spray intensity 20 %)
ES
M2 = (14.000 rpm, spray intensity 40 %)
M3 = (20.000 rpm, spray intensity 50 %)
M4 = (30.000 rpm, spray intensity 70 %)
PT
These settings may be changed as per description.
PL
of the rotation and function indicator (4) on
Kaitstud / Secured by https://mereneid.eu/
J. Direction of rotation, clockwise rotation / anti-
clockwise rotation
• The preset direction of rotation is clockwise rotation and is
shown on the display by the rotation and function indicator (4).
• By pressing the (C) button, the direction of rotation can be
changed.
• It is recommended to change the direction of rotation only
when the motor is at standstill. Changing the direction of
rotation at a high speed must always be avoided, as this
constitutes a very high load for the motor.
K. Spray setting
• The Promed 5040-SX2 is equipped with a high-
performance micropump generating finest spray.
• The spray volume is adjusted via the (H) or (I) button and can
be set in steps of 5 % using the (H)
• If you require a spray volume of 80 %, for example, press the
(H)
button until (F)
• The Promed 5040-SX2 offers the possibility of individually
adjusting the spray pattern of the spray to the spray liquid
used.
• To set the spray pattern, press and hold the (G)
for approx. 3 seconds. You will hear a beep and the scale for
setting the spray pattern is shown on the display.
or (I)
button.
is shown on the display.
button