PeratInG the UnIt
• The spray pattern of the spray can be increased (coarsened)
by pressing the (D)
pressing the (E)
button.
• Press the (G)
button as soon as you have set the required
spray to return to the "standard display".
L. SPRAY- Standby-Function
• In order to interrupt the supply of spray, press the button (G)
this „pause-Standby-function" represents the spray output. The
compressed air pump continues to operate so that the nozzle
blows out only the air. The spray supply is interrupted. During the
spray pause, the spray mode (6) indicator flashes on the display.
The spray intensity setting remains unchanged during the pause.
• By pressing the (G) button again, the spray supply is
reactivated and the spray mode (6) indicator is illuminated
on the display.
button or reduced (refined) by
Kaitstud / Secured by https://mereneid.eu/
Note: If you switch the spray function of the device
off, spray will still come out of the nozzle for a short period of
time until the hose has been emptied completely.
M. Overload protection for motor handpiece
• In case of any blockage or strain of the handpiece, the engine
and micropump turns off for approx. 5 seconds
• The handpiece overloading is shown on the display (F) by
an error message
disappears and the unit is in standby mode.
• If the handpiece is defective, this will be shown on the display
(F) by means of the error message
,
• In this case the complete device (control with handpiece) should
be sent to an authorized dealer for maintenance and repair.
N. Fuse
• The device fuse (P) is located on the rear side of the device.
• Before changing the fuse, always pull the mains plug out of
the socket.
• To replace the fuse, you must unscrew the fuse holder with
the screw driver and use fuse T 0,8A (for 115 volt version:
T 1,6A).