energy, and if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to radio or television reception, which
can be determined by turning the equipment off and on, the user is encouraged
to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Stacking: This device should not be used adjacent to or stacked with other
equipment. And if adjacent or stacked use is necessary, this device should be
observed to verify normal operation in the configuration in which it will be used.
Disposal: The symbol which can be found on the product or on the package
surface indicates that the product shall not be disposed with other household
wastes. It is the users responsiblity to properly dipsose of the product, or facilitate
recycling activities. For further information about the recycling collection points,
please contact your local authority or Lansinoh Customer Service.
EN 60601-1-2 STANDARD TEST REQUIREMENTS
Immunity Test Levels for ENCLOSURE PORT
Phenomenon
ELECTROSTATIC DISCHARGE
Radiated RF EM fields
Proximity fields from RF
wireless communications
equipment
RATED power frequency
magnetic fields
d) e)
26
Basic EMC
standard or
test method
IEC 61000-4-2
IEC 61000-4-3
a)
IEC 61000-4-3
IEC 61000-4-8
IMMUNITY TEST LEVELS
Professional healthcare
facility environment
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
3 V/m
f)
80 MHz – 2,7 GHz
b)
80 % AM at 1 kHz
c)
See 8.10
30 A/m
g)
50 Hz or 60 Hz
HOME HEALTHCARE
ENVIRONMENT
10 V/m
f)
80 MHz – 2,7 GHz
b)
80 % AM at 1 kHz
c)