Descargar Imprimir esta página

Lansinoh SMARTPUMP 2.0 Instrucciones De Uso página 27

Publicidad

Idiomas disponibles

Idiomas disponibles

Immunity Test Level for Input A.C. Power PORT
Phenomenon
Electrical fast
transients/bursts
a) l) o)
Surges
a) b) j) o)
Line-to-line
Surges
a) b) j) k) o)
Line-to-ground
Conducted
disturbances
induced by RF fields
c) d) o)
Voltage dips
f) p) r)
Voltage Interruptions
f) i) o) r)
Conducted Emission Test Level
Frequency range
MHz
0,15 – 0,50
0,50 – 5
5 – 30
At the transition frequency, the more stringent limit shall apply.
Radiated Emission Test Level
Frequency range
30 – 230
230 – 1 000
Basic EMC
standard or test
method
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Quasi-peak
dB(¨V/m)
66
Decreasing linearly with logarithm of
frequency to
56
56
60
Quasi-peak
dB(¨V/m)
40
47
IMMUNITY TEST LEVELS
Professional healthcare
facility environment
± 2 kV
100 kHz repetition frequency
± 0,5 kV, ± 1 kV
± 0,5 kV, ± 1 kV, ± 2 kV
3 V
m)
0, 15 MHz – 80 MHz
6 V
in ISM bands between
m)
0,15 MHz and 80 MHz
n)
80 % AM at 1 kHz
e)
0 % UT ; 0,5 cycle
g)
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT ; 1 cycle
and
70 % UT ; 25/30 cycles
h)
Single phase: at 0°
0 % UT ; 250/300 cycles
56
Decreasing linearly with logarithm of
frequency to
46
46
50
HOME HEALTHCARE
ENVIRONMENT
3 V
m)
0, 15 MHz – 80 MHz
6 V
in ISM and amatuer
m)
radio bands between 0,15
MHz and 80 MHz
80 % AM at 1 kHz
q)
h)
Average
dB(¨V/m)
n)
e)
27

Publicidad