Descargar Imprimir esta página

Boekel SERO 12 Manual De Instrucciones página 12

Publicidad

Idiomas disponibles

Idiomas disponibles

D
C
ELETING A
YCLE
Enter The advanced
Navigate to Delete
Confirm Deletion
C
L
YCLE
OCK
To ensure repeatability, the centrifuge can be locked either on one cycle (Single Cycle Lock) or restricted to the saved
cycles (Preset Lock). The Single Cycle Lock also prevents making changes to the selected cycle parameters. The Preset
Lock allows selection of any saved cycle and prevents changing the parameters of saved cycles.
Enter Preset Lock
Enter Single Cycle
Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 10
With the desired cycle selected, access the menu and enter the
menu
advanced menu.
Using the UP and DOWN buttons, navigate to DELETE. Exit the
menu. WARNING: CYCLE WILL BE DELETED IF MENU IS EXITED
WITH DELETE SELECTED
Press the CYCLE button to Delete the cycle
Select desired cycle. With lid open, press and hold the UNLOCK
button. One beep will confirm that cycle selection is locked.
Continue holding the UNLOCK button to enter Single Cycle Lock
Two beeps will confirm that cycle selection is now locked. NOTE:
If preset lock is set, it must be canceled before Single Cycle lock
can be set
Hold the UNLOCK button. Three beeps will confirm that the cycle
Cancel Lock
selection is now unlocked.

Publicidad

loading