Descargar Imprimir esta página

Boekel SERO 12 Manual De Instrucciones página 17

Publicidad

Idiomas disponibles

Idiomas disponibles

C
ALCULATING THE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which can be calculated if you know
the RPM and the radius. Use the formula below or go to www.druckerdiagnostics.com/g-force-calculator/.
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)
Radius
2.8 in (7.1 cm)
R
P
EPLACEMENT
ARTS
Part No.
Description
7724177K
Foot, rubber (Pack of 4)
02-002-1-0044
Lid Tray Assembly
420545
12 Place Rotor Assembly
420526
6 Place, 100mm Rotor Assembly
03-1-0006-0008
Motor Assembly
02-006-0-0021
PC Board
7760006
Power cord
03-1-0005-0193
Power Supply, 350W
02-002-1-0041
Lid Assembly
7724071K
Hinge, friction (Pack of 2)
02-002-1-0057
Seal, lid gasket
03-0-0003-0332
Open/Close Label
02-006-1-0066K
Lid LED PCBA Assembly, Red
03-0-0003-0823
Sero 12 Front Panel Label, Boekel
00-100-100-010
12/24 Series Soft Buttons & Spacer Replacement Kit
00-100-100-005
Replacement Grommet & Bushing Kit
Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 15
G-F
ORCE
2
In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)
2