The cycle time and speed are not set to
the desired value
The lid does not open
Clicking noise during braking gets loud
Lid does not stay up
G
S
ENERAL
PECIFICATIONS
The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
Tube Capacity
Dimensions (W x D x H)
Weight
Sound Level
Environmental Range
Voltage
Frequency
Power Requirement
Centrifuge Motor
Max G-Force
Max Speed
Min Cycle Time
Max Cycle Time
Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 14
o Check the setting by following the instructions in the section on
Changing Cycle Settings. If the preset is not the desired length,
follow the procedure on the same page to change the preset
time.
o Ensure that the lid knob is turned fully counterclockwise.
o If the knob cannot be turned counterclockwise, turn it fully
clockwise, press UNLOCK, and turn counterclockwise.
o If the lid remains locked after this and will not unlock, the
electronics may have been damaged. Contact customer service
for assistance.
o Make sure that the screw in the center of the rotor is tight.
o Tighten the center screw on the lid hinge.
6 capped tubes, 13 x 75 mm
12 uncapped tubes, 13 x 75 mm
13 in x 15 in x 9 in
(33 cm x 38 cm x 23 cm)
27 lbs. (12.3 kg)
61 dB A
16 – 32°C
100-240 VAC (± 10%)
50/60 Hz
415 Watts
½ H.P. Brushless
1,040 xg
3,600 RPM
15 seconds
99 minutes and 45 seconds