Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
Electrostatic dis‐
±8 kV contact
charge (ESD)
±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±15 kV air
Conducted distur‐
3 V
bances induced
150 kHz to 80 MHz
by RF fields
6 V
ISM bands between
150 kHz to 80 MHz
Electrical fast tran‐
±2 kV 100 kHz repetition fre‐
sient/burst
quency
IEC 61000-4-4
Surge
Line to Line: ±0.5 kV, ±1 kV
IEC 61000-4-5
Line to Ground: ±0.5 kV,
±1 kV, ±2 kV
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Voltage dips, short
0% U
interruptions, and
90°, 135°, 180°, 225°, 270°,
voltage variations
and 315°)
on power supply
0% U
lines
70% for 25/30 cycles (at 0°)
IEC 61000-4-11
0% for 250/300 cycles
Power frequency
30 A/m
(50/60 Hz) mag‐
netic field
IEC 61000-4-8
Proximity fields
IEC 60601-1-2
from RF wireless
communications
equipment
IEC 61000-4-3
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
IEC 60601-1-2 Test Level
RMS
RMS
; 0.5 cycles (at 0°, 45°,
T
; 1 cycle (at 0°)
T
Max foreseeable
use condition
per IEC
60601-1-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Not applicable
Not applicable
Not applicable
Not applicable
30 A/m
IEC 60601-1-2
10 V/m
80 MHz to 6 GHz
30
Electromagnetic Environ‐
ment Guidance
For use in a typical domestic,
commercial, or hospital envi‐
ronment.
Requirement does not apply
to this battery powered de‐
vice.
Requirement does not apply
to this battery powered de‐
vice.
Requirement does not apply
to this battery powered de‐
vice.
Requirement does not apply
to this battery powered de‐
vice.
For use in a typical domestic,
commercial, or hospital envi‐
ronment.
For use in a typical domestic,
commercial, or hospital envi‐
ronment.
Portable and mobile RF com‐
munications equipment