M i c r o C a r r i e r S t i r r e r
MCS-101L
•
Remove the two screws which holds the location spacer and
magnet carrier to the pulley.
•
Fit the appropriate magnet spacer to move the magnet position,
near the spindle for 3 litre vessels, away from the spindle for 5 litre
vessels. (The spacer and location boss for 3 litre vessels are
screwed inside the base of the platform).
•
Use the two screws to replace the spacer and carrier on the pulley.
•
Place the magnets so that they touch the spacer.
•
Note: Magnets must be mounted with opposite poles upwards
(opposite poles attract).
Spacer
Magnets
Other MCS Units
•
Remove the screw holding the
magnet to the pulley.
•
Position the magnet on the appropriate
peg as shown in the table.
Position
Vessel Size
Peg A
500 ml
Peg B
1 litre
Peg C
125/250 ml
•
Magnets must be mounted with their
blue spots (south poles) leading.
•
Replace the screw to hold the
magnet to the pulley.
O P E R A T O R ' S M A N U A L
Screws
Carrier
P A G E 19