Calibration Mode
GB
The result is shown in the display, alternating with CAL.
RESULT
If the reading corresponds with the value of the calibration standard
CAL
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Mode
Pressing the [ZERO/TEST] key once decreases the displayed value by
Zero
1 digit.
Test
CAL
Press the corresponding key until the reading equals the value of the
calibration standard.
RESULT + x
By pressing the [ON/OFF] key, the new correction factor is calculated
On
and stored in the user calibration software.
Off
Cal
:
:
Confirmation of calibration (3 seconds).
96
All manuals and user guides at all-guides.com
MD100_7 12/2016