Always use original components and supplies. Other devices can be similar
but they can damage the equipment.
Cleaning
•
Never use scourers or substances that can grate for cleaning metallic parts
such as stainless steel, aluminium, coatings, etc. as they damage the balance
and produce an early ageing of the equipment.
•
Use a fluff-free cloth dampened with soaped water that does not contain
abrasives.
ATTENTION!! IF EQUIPMENTS ARE NOT PROPERLY CLEAN AND DISINFECTED
THEY WOULD NOT BE ALLOWED TO REPAIR BY OUR TECHNICAL SERVICE.
Troubleshooting
Problem
Upper line - - - -
Under line _ _ _ _
Err-1
Err-2
Cause
Overload
Sample is not properly loaded on
the balance
Too frequently turn on and off the
balance
The balance is not stable
Low battery
21
Solution
Make sure the wait of the simple does not
exceed the maximum capacity of the
balance.
Re-calibrate the balance
Re-calibrate the balance
Turn off the balance, resume it after 3
seconds
Wait for a few seconds for stabilization
Replace batteries or use the balance
connected to the power supply